      Principle

      The airmoVOC C6-C12 uses a valve with 1 sample trap, and a metallic  capillary column.

      • Miniaturization, sensitivity, mobility and flexibility are its main features.

      • Uninterrupted sampling with pre-concentration on 1 absorbent tube

      • Gas chromatograph with 0.2 mm ID metallic column and programmable temperature gradient oven and pressure / flowcontrol of the  carrier gas by piezo-valve

      Before delivery the analyser is tested for one week by our quality control  department. The Vistachrom software enables the user to visualize and store 

      data on a PC. Furthermore it enables the user to recalculate, calibrate the instrument export data and to set-up measurement. 

      The software allows the calculation of retention time, area, mass or concentration profiles.

      Quantity :
      No Product available
      • Delivery policy Delivery time in Lithuania 1-2 working days, to other EU countries 3-5 w.d. to other non-EU 5-9 w.d.
      • Return policy Order could be returned in 15 days.

      Product technical specifications:

      C6 to C12 analysis:

      • C6 (Hexane, Dimethylbutane) to C10 

       (Diethylbenzene, Naphtalene) and dodecane

      Detection limit:

      • Tri-methylbenzene: 10 ppt or 0.05 µg/m3

      • Benzene: 10 ppt in standard

       1 ppt with airmoVOC Expert

      Detection range:

      • 0.05 to 45 µg/m3

       for Tri-methylbenzene and Benzene - (A3)

      • 0,50 to 400 µg/m3 for Tri-methylbenzene and Benzene - (A2)

      Note: Be careful with total mass and mass by compounds.

      Amplification 1: On demand to decrease the sampling volume.

      Relative Standard Deviation:

      • < 0.3 % over 48h (RT) 

      • < 3 % over 48 h (Conc.)

      Results:

      • Data storage

      • MODBUS / JBUS communication protocol (option)

      Cycle time:

      • 15 min, 30 min, 40 min, 60 min

      Gas supply:

      • H2 (FID and carrier gas): 30 ml/min (inlet 2 bars; 1/16’’  swagelock)

      • Air (FID): 180 ml/min (inlet 3 bars ; 1/8’’ swagelock)

      • Sample inlet (vacuum pump) 1/4’’ swagelock 

      • Pneumatic valve 90ml/commutation

      Sample volume:

      • 30 to 700 ml (programmable)

       (Sampling time up to 40 minutes)

      Power supply:

      • Main: 230 V / 115 V or 50 Hz / 60 Hz

      • Battery: 24 V (option)

      Electrical consumption:

      • Mean: 150 VA, Peak 360 VA

      Dimensions and weight:

      • Rack: 19’’ - (5U) 

      • Height: 222 mm

      • Width: 482 mm

      • Depth: 600 mm

      • Net Weight: 22 kg.