- Stability: 0.0075% of reading per year
- Time to setpoint: 20 seconds with no overshoot
- Control stability: 10 ppm
- Active matrix color screen with enhanced navigation menus
Standard Pressure Ranges
Model 7250xi
20, 50, 100, 300, 500, 1000 and 2500 psig (1.6, 4, 6, 25,
40, 60 and 160 barg)
Model 7250i
5, 10, 20, 50, 100, 300, 500 and 1000 psig (400 mbar, 1.0,
1.6, 4, 6, 25, 40 and 60 barg)
15, 20, 30 and 50 psia (1.0, 1.6, 2.5 and 4 bara)
Model 7250
5, 10, 20, 30, 50, 100, 150, 300, 500, 1000, 1500, 2000
and 2500 psig (400 mbar, 600 mbar, 1.0, 1.6, 2.5, 4, 6, 10,
16, 25, 40, 60, 100 and 160 barg)
15, 20, 30 and 50 psia (1.0, 1.6, 2.5, 4 and 210 bara)
Optional Pressure Ranges
Model 7250xi
Any full scale range from 20 to 2500 psig (1.6 to 172 barg)
Model 7250i
Any full scale range from 5 to 1000 psig
(400 mbar to 69 barg)
Model 7250
Any full scale range from 5 to 2500 psig (400 mbar to 172 barg) to 3000 psia (210 bara)
Optional Modes
Absolute using barometric reference sensor for ranges
from 15 to 2500 psig (1.0 to 172 barg)
Vacuum (negative gauge) for ranges 15 to 2500 psig (1.0 to 172 barg)
Vacuum reference (requires external vacuum pump)
Total Uncertainty
The maximum deviation from the true value of pressure
including precision, stability, temperature effects and the
calibration standard is:
Model 7250xi
5% to 100% FS
90 day 0.006% reading
1 year 0.009% reading
Model 7250i
25% to 100% FS
90 day 0.006% reading
1 year 0.009% reading
Model 7250
Ranges to 2500 psi (172 bar)
90 day 0.003% FS + 0.002% reading
1 year 0.003% FS + 0.0075% reading
3000 psi (210 bar)
90 day 0.01% of each range
1 year 0.014% of each range