- Temperature compensation from -20°C to 50°C
- Pressure accuracy to 0.025% of reading
- Weight: 680g.
- Dimensions: 178mm x 110mm x 58mm
- Field-replaceable modules: pressure, temperature, current and voltage
- ATEX/IECEx intrinsically safe
- Log and display 500,000 points at up to 10 readings/sec on up to two modules simultaneously
Backlit Display
- Description: 255 x 160 pixel graphical display with integrated backlight
- Display rate: selectable up to 10 readings per second
Power
- AA Battery Module: Four size AA (LR6) batteries; user serviceable, non-chargeable
- Battery life: 200 hours typical (alkaline battery)
- Note: Chassis can be powered remotely by the mini-USB connector on the side of the product
- WARNING: Do not use USB interface in a hazardous atmosphere
Temperature
- Operating & Compensated: -20 to 50°C (-4 to 122°F)1
- Storage: -40°C to 75°C (-40°F to 167°F)
1No change in accuracy over operating range for all modules. Pressure sensor must be zeroed to achieve rated specification
Humidity
- < 95% RH, non-condensing
IP Rating
- IP67 (1m immersion for 30 min) per ISO 60529
- Price:
This e-mail address is being protected from spambots. You need JavaScript enabled to view it.
-
This e-mail address is being protected from spambots. You need JavaScript enabled to view it.
|